Getting Started
VX1410 IntelliFrame Instruction Manual
1–17
WARNING.
Installing the VX1410 IntelliFrame in the rack requires at least two
people. Because of the size and weight of the mainframe, installing the main-
frame alone can result in personal injury or equipment damage. One person
should be on the left side of the mainframe with one hand holding the left front
handle and with the other hand underneath the mainframe. The other person
should do likewise on the right side of the mainframe.
11. Fully extend the inner tracks of each slide-out section out of the front of the
rack until they lock.
12. Lift the mainframe and insert the left and right tracks that extend from the
rear of the mainframe into the ends of the tracks that you just extended.
Make sure that the tracks mounted on the mainframe slip just inside the inner
tracks that you extended earlier.
13. Slide the mainframe back until it stops.
14. Push to release the button latches, located on the outside of each track, and
continue to slide the mainframe all the way back into the rack cabinet.
15. Tighten the four screws that you left loose at the rear of the rack when you
performed step 8 and pull the mainframe part way out of the rack. (Tighten
the 10-32 screws using 28 inch-lbs of torque.)
16. Be sure that the screws that you left loose at the front of the rack are loose
enough to allow the slide-out track assemblies to seek the normal positions.
17. Tighten the screws and push the mainframe all the way into the rack. If the
tracks do not slide smoothly, readjust the level using the method described in
steps 15 and 16. (Tighten the 10-32 screws using 28 inch-lbs of torque.)
18. Install the proper power cord at the rear of the mainframe.
Refer to Figure 1–9 and perform the following steps to install the rackmount
door for Option 2R:
1. Install the bottom halves of the two hinges to the left rackmount bracket as
shown in Figure 1–9.
2. Install the top halves of the hinges on the door.
3. Install the catch on the right rackmount bracket.
You can adjust the catch position forward or back to assure a flush fit
between the door and the rackmount bracket.
4. Lift the rackmount door and mate the door with the hinges.
Install the Option 2R
Rackmount Door