DISPlay Subsystem
3–28
VX1410 IntelliFrame Instruction Manual
DISPlay:WINDow:TEXT:DATA
DISPlay:WINDow:TEXT:DATA?
Sets or queries the user defined message to be displayed on the front panel
readout of the mainframe. You can display up to 20 characters in the message.
The string “SYSTEM OK” is the default factory message. This message only
displays when there are no monitor failures.
! "
!
Parameter
Query Response
<String>
<Block>
<String>
Not Applicable
–223, “Too Much Data”
You attempted to send too much data to the display.
–220, “Parameter Error”
You attempted to send nonprinting data in the block parameter.
None
Command:
#
Query:
Response:
#
None
NOTE
. (DISPlay:WINDow:TEXT:DATA): This command only applies to
mainframes with Option 1M. Please note that Option 2M does not include a
front panel display.
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com