STATus Subsystem
VX1410 IntelliFrame Instruction Manual
3–53
STATus:QUEStionable:TEMPerature?
Returns the contents of the Temperature Event register. A value of one in a bit
indicates the corresponding condition bit changed from false to true. Unused bits
always return a value of zero. The register is cleared by the *CLS command.
There is no command associated with this query.
Parameter
Query Response
Not applicable
<NR1>
Blower Status Word Bits
Definition
D0
Slot 0 Delta Temperature
D1
Slot 1 Delta Temperature
D2
Slot 2 Delta Temperature
D3
Slot 3 Delta Temperature
D4
Slot 4 Delta Temperature
D5
Slot 5 Delta Temperature
D6
Slot 6 Delta Temperature
D7
Slot 7 Delta Temperature
D8
Slot 8 Delta Temperature
D9
Slot 9 Delta Temperature
D10
Slot 10 Delta Temperature
D11
Slot 11 Delta Temperature
D12
Slot 12 Delta Temperature
D13
Ambient Temperature
D14
Not Used
D15
Not Used
Not Applicable
None