Status and Events
VX1410 IntelliFrame Instruction Manual
3–137
Table 3–20: VX1410 IntelliFrame Error and Event Codes
Code
Message
0
No error. The error queue is empty. Every error or event has been read or
was purposely cleared.
–100
Command error.
–101
Invalid character.
–102
Syntax error.
–103
Invalid separator. The parser detected an illegal separator character.
–104
Data type error. The parser recognized a data element different than the
one allowed.
–108
Parameter not allowed. More parameters than expected were received for
the header.
–109
Missing parameter. Not enough parameters were received.
–110
Command header error. An error was detected in the header.
–112
Program mnemonic too long. The header has more than 12 characters.
–113
Unidentified header The header is undefined for the VX1410 IntelliFrame.
–120
Numeric data error.
–141
Invalid character data. The character data element contains an invalid
character or the data was invalid for the header.
–144
Character data too long. The character element was more than 12
characters long.
–160
Block data error.
–161
Invalid block data. An invalid block data element was received.
–200
Execution Error.
–203
Command protected. A password-protection program command or query
could not be executed because the command was disabled.
–213
Init Ignored. The measurement initiation was ignored because another
measurement was already in progress.
–220
Parameter error. A program data element related error occurred.
–221
Settings conflict. A legal program data element was parsed but could not be
executed due to the current state of the mainframe.
–222
Data out of range. The value of the data was outside of the legal range.
–223
Too much data. More data was received by the instrument than what it was
capable of handling.
–224
Illegal parameter value. The parameter value must be one of the exact
values from the command list.
–315
Queue Overflow. An error occurred but there was no room in the queue to
record the error.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com