STATus Subsystem
3–38
VX1410 IntelliFrame Instruction Manual
STATus:OPERation:VXIMonitor:ENABle
STATus:OPERation:VXIMonitor:ENABle?
Sets or queries the contents of the VXI Monitor Enable register. Setting unused
enable bits does not generate an error; the bits are ignored.
If the power-on status clear flag is FALSE (see *PSC command on page 3–124)
at power-on, the register value will be restored from nonvolatile RAM; otherwise
the register value at power-on is set to #H7FFF. Executing the STATus:PRESet
command sets the register value to #H7FFF.
STATus:OPERation:VXIMonitor:ENABle <Mask>
STATus:OPERation:VXIMonitor:ENABle?
<Mask>
Query Response
{<NRf> | <Nondecimal Numeric>}
0
≤
N
≤
#HFFFF
<NR1>
Not Applicable
–222, “Data Out of Range”
You attempted to set the enable register to an illegal value.
None
Command:
STAT:OPER:VIXM:ENAB #H0001
Query:
STAT:OPER:VXIM:ENAB?
Response:
1
STATus:OPERation:VXIMonitor?
STATus:PRESet
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands