Status and Events
3–136
VX1410 IntelliFrame Instruction Manual
Error and Event Messages
Table 3–20 on page 3–137 lists some of the error messages that you may
encounter while sending commands and queries to the VX1410 IntelliFrame.
The VX1410 IntelliFrame reports the following categories of errors:
H
Command Errors
H
Execution Errors
H
Device Specific Errors
H
Query Errors
Command errors have error codes within the range of –100 and –199. These
errors indicate syntax errors. The syntax errors can be due to the command parser
detecting an IEEE 488.2 syntax error or an unrecognized header was received.
Command errors may also occur when you attempt to send an unimplemented
IEEE 488.2 common command.
Execution errors have error codes within the range of –200 and –299. These
errors are detected by the VX1410 IntelliFrame execution control block.
Execution errors can occur when a program data element following a header was
evaluated as outside of the legal input range or when the program data element is
inconsistent with the capabilities of the mainframe.
Device specific errors have error codes within the range of –300 and –399. These
errors may indicate that some operations failed to complete due to abnormal
hardware or firmware conditions.
Query errors have error codes within the range of –400 and –499. These errors
indicate that the VX1410 IntelliFrame detected a problem with the message
exchange protocol. Query errors set the query error bit in the Event Status
register. These errors may result when you attempt to read data from the output
queue when no input is present or pending. The errors may also occur when the
data in the output queue is lost.
Command Errors
Execution Errors
Device Specific Errors
Query Errors