TEST Subsystem
VX1410 IntelliFrame Instruction Manual
3–111
TEST
TEST?
This command executes all internal self-tests. If a failure occurs and the halt
control is enabled, the test will immediately end. The query form returns the
unique numeric identifier of the first test that failed. A value of zero is returned if
there are no failures. The command form executes the same tests but returns no
results code.
Each internal self-test has a unique numeric identifier in the range of 1000 to 1999
(see Table 3–19 on page 3–109).
*TST? and TEST? are aliases.
Parameter
Query Response
Not Applicable
<NR1>
0
1000
≤
N
≤
1999
Not Applicable
None
None
Command:
Query:
Response:
TEST:RESults?
TEST:RESults:VERBose?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com