Adjustment Procedures
6–2
VX1410 IntelliFrame Instruction Manual
7. Connect an RS-232 cable between the COM port on the PC and the RS-232
connector at the rear of the VX1410 IntelliFrame. Refer to Table 3–14 on
page 3–77 for a list of the factory default RS-232 settings.
8. Select the Communications... command from the Settings menu of the
Terminal program and select the appropriate COM port (for example, COM1
or COM2) from the dialog box.
9. Refer to Table 6–1 and set up the PC RS-232 port configurations. The
RS-232 settings must match those of the mainframe.
Table 6–1: PC RS-232 Settings for the Calibration
Procedure
RS-232 Parameter
Setting
Baud Rate
9600
Data Bits
8
Stop Bits
1
Parity
None
Flow Control
Xon/Xoff
10. Make sure that the mainframe has had at least five minutes to warm up
before continuing this procedure.
11. Send the following commands to enable the password and to calibrate the
VX1410 IntelliFrame:
12. Check that the CAL? query returns a value of zero to indicate no error has
been detected during the calibration.
13. Note the serial number on the rear panel of the mainframe.
14. Send the following command to set up the serial number in the enhanced
monitor:
The value of Bxxxxxx is the serial number as noted from the rear of the
mainframe.
15. Use the front panel Next and Prev buttons to scroll through the delta
temperature of the slots. All delta temperatures should read 0
°
C
±
1
°
C.
16. Power down the mainframe.