Maintenance
7–12
VX1410 IntelliFrame Instruction Manual
To remove the Temperature Sense board, remove the mainframe cover and then
complete the following steps:
1. Disconnect the ribbon cable from the Temperature Sense board.
2. Refer to Figure 7–9 and gently pry back on each retainer holding the
Temperature Sense board in place.
3. Lift the board out of the holes at the top of the mainframe being careful not
to damage any components on the circuit board.
NOTE
. If you replace the Temperature Sense board and you have Option 1M or
2M, remember to calibrate the mainframe. Refer to Chapter 5 Adjustment
Procedures beginning on page 6–1 for information on calibrating the main-
frame.
Temperature
Sense Board
Cable
Retainer
Figure 7–9: Removing the Temperature Sense Board
Procedure 6: Remove the
Temperature Sense Board