Maintenance
VX1410 IntelliFrame Instruction Manual
7–3
CAUTION.
Avoid getting moisture inside the mainframe during exterior cleaning;
use just enough moisture to dampen the cloth or swab.
Do not wash the front-panel On/Standby switch. Cover the switch while washing
the mainframe.
Use only deionized water when cleaning. Use a 75% isopropyl alcohol solution
as a cleanser and rinse with deionized water.
Do not use chemical cleaning agents; they may damage the mainframe. Avoid
chemicals that contain benzene, toluene, xylene, acetone, or similar solvents.
The fan filter easily removes from the rear of the mainframe. The filter slides out
of the rear of the mainframe as shown in Figure 7–1.
Fan Filter
Figure 7–1: Removing the Fan Filter
Clean the fan filter by vacuuming or blowing air through the filter. Rinse the
filter in cool water and dry it before replacing it in the mainframe.
If your mainframe has the enhanced monitor (Option 1M or 2M), you can reset
the filter cleaning interval with the VXIplug&plug soft front panel, or by sending
the STATus: QUEStionable: FILTer:TRESet command. If you mainframe has the
front panel display (Option 1M only), you can also reset the filter cleaning
interval by scrolling to the air filter cleaning time and simultaneously pressing
both the NEXT and PREV buttons.
Cleaning the Fan Filter