STATus Subsystem
VX1410 IntelliFrame Instruction Manual
3–59
STATus:QUEStionable:TEMPerature:LIMit
STATus:QUEStionable:TEMPerature:LIMit?
Sets or queries the maximum allowable ambient and delta temperatures in
each slot. The factory default settings are 55
°
C for ambient and 10
°
C for each
delta temperature.
<Location>
Query Response
{AMBient | DELTa0 . . . DELTa12}
AMBient – ambient input temperature
DELTa0 . . . DELTa12 – output temperature minus
input temperature of each slot
<NR1>
<Limit>
<NRf>
If Location is AMBient
1
≤
Limit
≤
55
MIN
MAX
If Location is DELTan
5
≤
Limit
≤
30
MIN
MAX
1
55
5
30
Not Applicable
–224, “Illegal Parameter Value”
You attempted to set the location to an illegal value.
–222, “Data Out of Range”
You attempted to set the temperature limit to an illegal value.
None