Specifications
4–6
VX1410 IntelliFrame Instruction Manual
Table 4–4: Safety
1, 2
Characteristic
Description
Safety Certification
The VX1410 IntelliFrame complies with the requirements
of UL3111-1, IEC 1010-1, and CAN/CSA–22.2 No 1010.1.
EN 61010–1 Safety requirements for electrical equipment
for measurement, control and laboratory use.
Does not include mainframes shipped as Option 1D,
Delete Power Supply.
IEC Characteristics
Installation Category II
Pollution Degree 2
Safety Class I
1
If the VX1410 IntelliFrame is configured with Option 1D, Delete Power Supply, all
product specifications in this table do not apply. To be VXI compliant, the power
supply must adhere to the specifications of the VXI standard.
2
If the VX1410 IntelliFrame is configured with Option 1D, Delete Power Supply, to
maintain safe operation, the power supply that provides power to the chassis should
not exceed the I
MP
current ratings found under DC Current Capacity (I
MP
) in
Table 4–2 on page 4–2. The short circuit capacity of each voltage should not exceed
120% of the I
MP
rating. The total power dissipation of the chassis, with modules
installed, should not exceed 925 Watts.