STATus Subsystem
VX1410 IntelliFrame Instruction Manual
3–49
STATus:QUEStionable:ENABle
STATus:QUEStionable:ENABle?
Sets or queries the contents of the Questionable Enable register. Setting unused
enable bits does not generate an error; the bits are ignored.
If the power-on status clear flag is FALSE (see *PSC command on page 3–124)
at power-on, the register value will be restored from nonvolatile RAM; otherwise
the register value at power-on is set to #H0000. Executing the STATus:PRESet
command sets the register value to #H0000.
<Mask>
Query Response
{<NRf> | <Nondecimal Numeric>}
0
≤
N
≤
#HFFFF
<NR1>
Not Applicable
–222, “Data Out of Range”
The value of the data was outside of the legal value.
None
Command:
Query:
Response:
STATus:QUEStionable?
STATus:QUEStionable:CONDition?
STATus:PRESet
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com