STATus Subsystem
VX1410 IntelliFrame Instruction Manual
3–75
STATus:SEVent?
Returns the summary information on the power supply voltage events, tempera-
ture events, blower events, filter events, and VXI monitor events.
A one in a bit indicates the corresponding event has been asserted. Unused bits
always return a value of zero. The query returns a value of zero at power on or
when you send the *CLS command. Reading events by this command does not
clear the events.
There is no command associated with this query.
Parameter
Query Response
Not applicable
<NR1>
Event Status Word Bits
Definition
D0
Blower 1
D1
Blower 2
D2
Filter
D3
Slot 0 Delta Temperature
D4
Slot 1 Delta Temperature
D5
Slot 2 Delta Temperature
D6
Slot 3 Delta Temperature
D7
Slot 4 Delta Temperature
D8
Slot 5 Delta Temperature
D9
Slot 6 Delta Temperature
D10
Slot 7 Delta Temperature
D11
Slot 8 Delta Temperature
D12
Slot 9 Delta Temperature
D13
Slot 10 Delta Temperature
D14
Slot 11 Delta Temperature
D15
Slot 12 Delta Temperature
D16
Ambient Temperature
D17
+24 V
D18
+12 V
Syntax
Parameters