STATus Subsystem
VX1410 IntelliFrame Instruction Manual
3–33
STATus:OPERation:ENABle
STATus:OPERation:ENABle?
Sets or queries the contents of the Operation Event Enable register. Setting
unused enable bits does not generate an error; the unused bits are ignored.
If the power-on status clear flag is FALSE (see *PSC command on page 3–124)
at power-on, the register value will be restored from nonvolatile RAM; otherwise
the register value at power-on is set to #H0000. Executing the STATus:PRESet
command sets the register value to #H0000.
STATus:OPERation:ENABle <Mask>
STATus:OPERation:ENABle?
<Mask>
Query Response
{<NRf> | <Nondecimal Numeric>}
0
≤
N
≤
#HFFFF
<NR1>
Not Applicable
–222, “Data Out of Range”
You attempted to set the mask to an illegal value.
None
Command:
STAT:OPER:ENAB #H0100
Query:
STAT:OPER:ENAB?
Response:
256
STATus:OPERation?
STATus:OPERation:CONDItion
STATus:PRESet
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands