Status and Events
VX1410 IntelliFrame Instruction Manual
3–133
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
VXI Monitor Status Registers
Not Used
BERR occurred
AS Missing
SYSFail occurred
ACFail occurred
STATus:OPERation:VXIMonitor[:EVENt]?
STATus:OPERation:VXIMonitor:CONDition?
STATus:OPERation:VXIMonitor:ENABle
Summary of VXI Monitor
Figure 3–12: VXI Monitor Status Register
The Questionable Status register contains information on the quality of the
signals within the mainframe. In the event problems occur within the mainframe
(for example, one of the power supplies is not working properly), you could
program the mainframe to warn you of possible problems.
There are three status registers that provide summary information to the
Questionable Status register: the Voltage Status register, the Temperature Status
register, and the Blower Status register.
The Voltage Status register provides status information on the individual power
supply voltages (see Figure 3–13). You can detect problems with any one of the
power supply voltages by monitoring the corresponding bit in the Voltage Status
register. If a problem occurs, the corresponding bit will be set in the register.