VX1410 IntelliFrame Instruction Manual
4–1
Specifications
This chapter contains the complete specifications for the VX1410 IntelliFrame.
Within each section, the specifications are arranged in functional groups such as:
AC Power Source, Secondary Power, Cooling, Safety, Environmental, Back-
plane, and Mechanical.
All specifications are warranted unless they are designated typical. Warranted
characteristics are directly checked by procedures contained in the Performance
Verification chapter beginning on page 5–1; these characteristics are listed in
boldface under the Characteristics column.
If the characteristic is noted as typical, the characteristic is not warranted.
Typical characteristics describe typical or average performance and provide
useful reference information.
Table 4–1: AC Power Source
1
Characteristic
Description
Source Voltage
90 V
RMS
to 250 V
RMS
, continuous range
Source Frequency
45 Hz to 440 Hz, continuous range, 115 V operation
45 Hz to 66 Hz, 230 V operation
Power Consumption
1350 W line power maximum
Fuse Rating
90 V – 132 V Operation
0.25 in
×
1.25 in, Slow Blow, 20 A, 250 V
103 V – 250 V Operation
0.25 in
×
1.25 in, Fast Blow, 15 A, 250 V
207 V – 250 V Operation
5 mm
×
20 mm, Fast Blow, 6.3 A, 250 V
Inrush Surge Current
70 A maximum
Input Current
15 A maximum at 90 V
RMS
6.3 A maximum at 207 V
RMS
Power Factor Correction
Yes
Power Disconnect
Front Panel On/Standby. Rear panel connection allows
for remote operation. Power cord provides main power
disconnect. The On/Standby switch can be disabled for
remote operation. With Option 1M, the mainframe can be
powered on or off under program control.
1
If the VX1410 IntelliFrame is configured with Option 1D, Delete Power Supply, all
product specifications in this table do not apply. To be VXI compliant, the power
supply must adhere to the specifications of the VXI standard.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com