7
GasLab Q2 displays and operating the device
Page 54 of
135
G a s Q u a l i t y A n a l y z e r
GasLab Q2
7.2.2 Display Info Q2
The following display is behind this icon :
The display shows the serial number of the
device and the status of the security switch
(SSW), a sealable (official) hardware
interrupt. This safety prevents illegal
changes to fiscal and imprortant device
settings, or software downloads. See
chapter 7.3.10.
Apart from that, you are also able to branch
to the following functions:
Figure 7.8: Display Info Q2
The
Device monitor
displays information
about operating hours and Battery capacity as
well as CPU- and RAM-load. The battery is
mainly is used or power is drawn from the
battery when the device is turned off. After
repairs or servicing (with battery change), a
user with appropriate permissions can reset
the battery status to 100%.
Figure 7.9: Display Info Device monitor
The battery (Li/SOCl
2
½ AA Voltage 3.6 V) must be replaced when the battery
capacity is less than 20%. The exchange is performed by Elster authorized
personnel, because the housing must be opened.
An empty battery may sometimes lead to a loss of the measurement data that
was calculated and saved. The device will then not start correctly without
support from Elster GmbH.