Technical data and notes
10
GasLab Q2
G a s Q u a l i t y A n a l y z e r
Page 119 of 135
10
Technical data and notes
The essential technical data is shown on the various type plates and labels directly on the device and are
reproduced here below. Supplementary tables and the CE Declaration complete this overview.
10.1 Tables and information additional to the device
Q2 System
Values and / or Comment
Location of use
Installation close to the sampling point up to Zone 1
Mounting clearance
left and right at least 300 mm / top and bottom at least 200 mm / free space
(recommended 900 x 900 mm)
Gas group
only gases which are according to appropriate gases below
IIC / (Group: A B C D pending)
It is not allowed to supply gasses which can react in the abcence of air!
Gas ports */
Gas connection
Input (1 x Process gas. 1 x calibration gas mixture)
Output (1 x Process gas bypass 1 x Measurement gas)
Swagelok fittings 6 mm or ¼ " could be supplied by Elster
Breather vent line
Breather connection: 1/4 "NPT
Gas management
Integrated 2-channel double block and bleed device
for process gas and calibration gas.
Size of housing /dimensions
H = 38 cm, W = 27 cm, D = 23 cm
Weight
Approximately 16kg
IP rating
IP64
Voltage / Curent / Power
24 V DC Fluctuation, including mains fluctuation ±15% / 4 A / 96 W
Operation panel
Color display 9,5 x 10,5 cm (illuminated); 7 key captive touchpad
2 multi-colored LEDs for indicating the power supply and status
Heat-up time
At -25 ° C up to 1 h
Housing heater
30 W
Dynamics
Analysis rate f=1 Hz, reaction time t90≤6s
(90% step response / Tested @ NMi)
Operational calibration interval
settin
1 Week
Service calibration interval
1 Year
Data memory (non-volatile):
Archives / Calibration data / Parameterization / Logbooks
Operating system
Service PC: Refer to the download area of Elster
Software
enSuite
File Format:
enSuite.csv / enSuite.par
Data logger
Integrated (Logger interval adjustable)
*Standard equipment other connections available on request