GasLab Q2 displays and operating the device
7
GasLab Q2
G a s Q u a l i t y A n a l y z e r
Page 75 of 135
Working steps Verification manual
3.
On the adjacent display, activate "
Start
"
(If step 2 has been executed you have to
activate
Verification
again)
Figure 7.49:
Display Q2 Verification
4.
The ongoing measurement (analysis) is
stopped, you will see the following display:
Figure 7.50:
Display Q2 Confirm (verification gas)
5.
Change the gas at the process input (PG) by connecting the cylinder with the test gas
through a pressure regulator according to chapter 5.3.4. Open the test gas cylinder; the
cylinder pressure regulator must be adjusted between 0.15 and 0.5 MPa (g).
Follow all safety notes when working on gas connections!
6.
After successful connection of the gas, completing the safety query from step 4, accept
with “
Confirm
” (See figure 7.50). Or you can cancel the process now with “
Cancel
”
7.
After the confirmation the device is
flushed with the test gas, then the
measurement will starting. (See figure).
Because there is too much data to show
at once, you can scroll with the overlay
keys. The information of the last or current
measurement is in the lower part of the
display and will be shown below the
"
Start
" or "
Stop
" link.
Figure 7.51:
Display Q2 Verification is running
8.
The verification is now running for the
specified time in enSuite (see chapter 8).
By selecting "
Stop
" you have the possi-
bility to end the verifycation prematurely,
you will direct jump to step 9.
After the time has elapsed, you will be
prompted to return to the analysis mode.
Select "
Start analysis
"
Figure 7.52:
Display Q2 Verification end
9.
You will be asked to restore the original
gas connection.
Figure 7.53:
Display Q2 Confirm (process gas)
10.
Professionally rebuild the original connection again.
Follow all safety notes when working on gas connections!
Perform a leak test!
11.
Accept after successful connection of the gas the safety query from step 9 (see figure
above) with "
Confirm
". The device returns to the analysis mode and the verification is
finished.