Parameter reference
8.6 Setup (2.)
HydroRanger 200 HMI
166
Operating Instructions, 06/2018, A5E36281317-AC
8.6.1.9
Short shot duration (2.1.9.)
Adjusts the duration of the short shot transmit pulse, in milliseconds.
Index
Single-point model
Dual-point model
Global
Transducer
Values
Range: 0.000 to 5.000
Altered by
Transducer (2.1.5.) (Page 164)
Related
•
Number of long shots (2.1.12.) (Page 167)
•
Long shot duration (2.1.10.) (Page 166)
•
Long shot frequency (2.1.8.) (Page 165)
8.6.1.10
Long shot duration (2.1.10.)
Adjusts the duration of the long shot transmit pulse, in milliseconds .
Index
Single-point model
Dual-point model
Global
Transducer
Values
Range: 0.000 to 5.000
Altered by
Transducer (2.1.5.) (Page 164)
Related
•
Number of short shots (2.1.11.) (Page 166)
•
Short shot frequency (2.1.7.) (Page 165)
•
Short shot duration (2.1.9.) (Page 166)
•
Short shot bias (2.11.2.10.) (Page 225)
•
Short shot floor (2.11.2.11.) (Page 226)
•
Short shot range (2.11.2.12.) (Page 226)
8.6.1.11
Number of short shots (2.1.11.)
Sets the number of short shots to be fired (and results averaged) per transmit pulse.
Index
Single-point model
Dual-point model
Global
Transducer
Values
Range: 0 to 100
Preset: 1
Related
•
Number of long shots (2.1.12.) (Page 167)
•
Short shot frequency (2.1.7.) (Page 165)
•
Long shot duration (2.1.10.) (Page 166)
•
Short shot bias (2.11.2.10.) (Page 225)
•
Short shot floor (2.11.2.11.) (Page 226)
•