Trigger
R&S
®
RTM3000
92
User Manual 1335.9090.02 ─ 08
"Equal"
Triggers on transition times inside the time range
Rise Time
±
Varia-
tion
.
"Not equal"
Triggers on transition times outside the time range
Rise Time
±
Varia-
tion
.
Remote command:
Rise Time
Sets the reference rise time, the nominal value for comparisons.
Remote command:
Variation
Defines a time range around the given "Rise Time" value.
Remote command:
Upper Level
Sets the upper voltage threshold. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected polarity.
Remote command:
TRIGger:A:LEVel<n>:RISetime:UPPer
Lower Level
Sets the lower voltage threshold. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected slope.
The value corresponds to the threshold value of the trigger channel.
Remote command:
TRIGger:A:LEVel<n>:RISetime:LOWer
Hysteresis
Hysteresis of the trigger source channel, see
Remote command:
CHANnel<m>:THReshold:HYSTeresis
Find Threshold
5.11
Timeout Trigger
The timeout trigger checks if the signal stays above or below the threshold voltage for
a specified time lapse. In other words, the trigger occurs if the trigger source signal
does not cross the threshold during the specified time.
Timeout Trigger