Measurements
R&S
®
RTM3000
160
User Manual 1335.9090.02 ─ 08
Coupling
If enabled, the cursors lines are coupled and moved together.
Press the [Navigation] key to select whether both cursors or one cursor is moved. If
coupling is disabled, pressing the [Navigation] key toggles the single cursor lines.
Remote command:
Set To Trace
Autoset for cursor lines, sets the cursor lines to typical points of the waveform depend-
ing on the selected cursor type. For example, for voltage measurement ("Horizontal"),
the cursor lines are set to the upper and lower peaks of the waveform. For time mea-
surement ("Vertical"), the cursor lines are set to the edges of two consecutive positive
or two consecutive negative pulses.
Remote command:
Cursor Measurements
www.allice.de
Allice Messtechnik GmbH