Remote Control Commands
R&S
®
RTM3000
492
User Manual 1335.9090.02 ─ 08
OR
At least one of the channels must have the required state.
NAND
"Not and" operator, at least one of the channels does not have
the required state.
NOR
"Not or" operator, none of the channels has the required state.
*RST:
AND
SEARch:TRIGger:PATTern:LEVel<n>
<ThresholdLevel>
Sets the threshold value for the specified source channel. You can set different levels
for the channels
Suffix:
<n>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<ThresholdLevel>
Range:
Depends on vertical scale
SEARch:TRIGger:PATTern:LEVel<n>:DELTa
<LevelDelta>
Sets a hysteresis range to the threshold of the specified source channel to avoid
unwanted search results caused by noise oscillation around the level.
Suffix:
<n>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<LevelDelta>
Range:
Lower limit depends on vertical scale and other set-
tings, no upper limit
SEARch:TRIGger:PATTern:WIDTh:RANGe
<Range>
Sets the condition how the duration of a steady pattern is compared with the given ref-
erence time.
To set the reference value
width
SEARch:TRIGger:PATTern:WIDTh[:WIDTh]
.
To set a range
Δt
, use
SEARch:TRIGger:PATTern:WIDTh:DELTa
Parameters:
<Range>
WITHin | OUTSide | LONGer | SHORter
WITHin
Finds patterns steady for a time range
width ± Δt
.
OUTSide
Finds patterns outside a time range
width ± Δt
.
Waveform Analysis
www.allice.de
Allice Messtechnik GmbH