Remote Control Commands
R&S
®
RTM3000
783
User Manual 1335.9090.02 ─ 08
PGENerator:PATTern:FREQuency
<PatternFrequency>
Defines the frequency (period) value of the pattern generator function.
Parameters:
<PatternFrequency> Numeric value
Default unit: Hz
16.14.2.3
Counter Pattern
PGENerator:PATTern:COUNter:FREQuency
.....................................................................783
PGENerator:PATTern:COUNter:DIRection
........................................................................ 783
PGENerator:PATTern:COUNter:FREQuency
<Period>
Defines the frequency value of the pattern generator counter function. The user fre-
quency always refers to the switching of the pattern condition. This results in square
waveforms for individual pins.
Pin
Frequency
S0
f/2
S1
f/4
S2
f/8
S3
f/16
Parameters:
<Period>
Numeric value
Range:
2.380952425301E-02 to 2.50000007
*RST:
1.00000005
Default unit: Hz
PGENerator:PATTern:COUNter:DIRection
<CountDirection>
Sets the pattern generator counter direction.
Parameters:
<CountDirection>
UPWard | DOWNward
*RST:
UPW
16.14.2.4
Arbitrary Pattern
PGENerator:PATTern:TRIGger:MODE
..............................................................................784
PGENerator:PATTern:ARBitrary:DATA[:SET]
.....................................................................784
PGENerator:PATTern:ARBitrary:DATA:LENGth
................................................................. 784
PGENerator:PATTern:ARBitrary:DATA:APPend
................................................................. 784
PGENerator:PATTern:ARBitrary:DATA:APPend:BOR
......................................................... 785
PGENerator:PATTern:ARBitrary:DATA:APPend:BAND
....................................................... 785
Signal Generation (Option R&S
RTM-B6)
www.allice.de
Allice Messtechnik GmbH