Remote Control Commands
R&S
®
RTM3000
527
User Manual 1335.9090.02 ─ 08
Parameters:
<Xposition1>
<Xposition2>
Range:
Depends on the current instrument settings, for
example, horizontal position.
Default unit: s
CURSor<m>:Y1Position
<Yposition1>
CURSor<m>:Y2Position
<Yposition2>
The commands specify or return the positions of horizontal cursor lines on the y-axis
(voltage, current, level for FFT).
Suffix:
<m>
.
1, the suffix is irrelevant.
Parameters:
<Yposition2>
Range:
Depends on the current instrument settings.
Increment: 0.01
Default unit: V
CURSor<m>:YCOupling
<Coupling>
CURSor<m>:XCOupling
<Coupling>
If enabled, the cursor lines are coupled so that the distance between the two lines
remains the same if one cursor is moved.
Suffix:
<m>
.
1, the suffix is irrelevant.
Parameters:
<Coupling>
ON | OFF
*RST:
OFF
CURSor<m>:SWAVe
Autoset for cursor lines, sets the cursor lines to typical points of the waveform depend-
ing on the selected cursor type. For example, for voltage measurement ("Horizontal"),
the cursor lines are set to the upper and lower peaks of the waveform. For time mea-
surement ("Vertical"), the cursor lines are set to the edges of two consecutive positive
or two consecutive negative pulses.
Suffix:
<m>
.
1, the suffix is irrelevant.
Usage:
Event
CURSor<m>:TRACking:SCALe[:STATe]
<State>
Enables the adjustment of cursor lines if the vertical or horizontal scales are changed.
Measurements