Remote Control Commands
R&S
®
RTM3000
453
User Manual 1335.9090.02 ─ 08
Parameters:
<AutomaticRoll>
ON | OFF
*RST:
OFF
TIMebase:ROLL:MTIMe
<MinTimeBase>
The roll mode is enabled automatically if the time base exceeds the
MinTimeBase
,
and if
Parameters:
<MinTimeBase>
Limit value for roll mode enabling.
*RST:
500e-3
Default unit: s/div
ACQuire:INTerpolate
<InterpolationType>
Defines the interpolation mode.
See also:
Parameters:
<InterpolationType> SINX | LINear | SMHD
LINear
Linear interpolation between two adjacent sample points.
SINX
Interpolation by means of a sin(x)/x curve.
SMHD
Sample & hold causes a histogram-like interpolation.
*RST:
SINX
ACQuire:POINts:ARATe?
Retrieves the sample rate of the ADC, that is the number of points that are sampled by
the ADC in one second.
Return values:
<ACDsampleRate>
ADC sample rate
Default unit: Hz
Usage:
Query only
ACQuire:SRATe?
<SampleRate>
Returns the sample rate, that is the number of recorded waveform samples per sec-
ond.
Parameters:
<SampleRate>
Default unit: Sa/s
Usage:
Query only