General Instrument Setup
R&S
®
RTM3000
227
User Manual 1335.9090.02 ─ 08
► To remove all waveforms and results from the display, press the [Clear Screen] key
at the front panel.
Persistence
Defines the persistence (afterglow effect) of the waveform on the display.
"Off"
Deactivates persistence.
"Manual"
User-defined persistence according to "Time" setting.
"Infinite"
Activates persistence with infinite duration. Each new data point
remains on the screen infinitely until this setting is changed or the
persistence is cleared
Remote command:
Time ← Persistence
Specifies the afterglow duration if "Persistence" is set to "Manual". Each new data
point remains on the screen for the time defined here. Values range from 50
ms (=
default) to 12.8
s.
Remote command:
Intensities
Provides functions to define the brightness (relative luminous intensity) of display ele-
ments and control LEDs.
You can also open this menu directly using the [Intensity] key.
Display Settings