Applications
All available applications are provided in the "Apps Selection" dialog.
► To select an application, press the
[Apps Selection] key.
See also:
Chapter 3.2, "Selecting the Application"
Applications are grouped on several tabs:
●
Applications, see below
●
Protocol
Protocol applications are described in
Chapter 12, "Serial Bus Analysis"
●
Track
On the "Applications" tab, the following applications are available:
●
"QuickMeas": see
Chapter 7.1, "Quick Measurements"
●
"Probe Adjust": see
Chapter 4.4.1, "Adjusting Passive Probes"
●
Chapter 15.1, "Function Generator"
●
"Pattern Gen.": see
Chapter 15.2, "Pattern Generator"
Spectrum Analysis and Spectrogram (Option R&S
............................. 177
Spectrum Analysis and Spectrogram (Option R&S
............................. 187
........................................................................202
8.1
Mask Testing
Masks are used to determine whether the amplitude of a signal remains within speci-
fied limits, e.g. to detect errors or test compliance of digital signals.
8.1.1
About Masks and Mask Testing
Masks
A mask is specified by an upper and a lower limit line. The signal must run inside these
limit lines, otherwise a mask violation occurs.
A new mask is created from an existing signal: Mask limits are created by copying the
envelope waveform, and the limits are moved and stretched. The result is a tolerance
tube around the signal that is used as mask.
Mask Testing