Remote Control Commands
R&S
®
RTM3000
461
User Manual 1335.9090.02 ─ 08
TRIGger:A:WIDTh:RANGe
<RangeMode>
Defines how the measured pulse width is compared with the given limits.
Parameters:
<RangeMode>
WITHin | OUTSide | SHORter | LONGer
WITHin | OUTSide
Triggers on pulses inside or outside a range defined by
time ±
delta
. The time is specified with
the range around is defined with
To trigger on an exact value ("Width
=" or "Width
≠
", set the
range to 0 ("Variation",
).
SHORter | LONGer
Triggers on pulses shorter or longer than a time set with
.
*RST:
LONGer
TRIGger:A:WIDTh:WIDTh
<Time1>
For the ranges WITHin and OUTSide (defined using
<Time1> defines the center of a range which is defined by the limits
±
<Delta> (set with
).
For the ranges SHORter and LONGer, the width defines the maximum and minimum
pulse width, respectively.
Parameters:
<Time1>
Center value, maximum value or minimum value depending on
the defined range type.
Range:
20E-9 to 6.87194685440
Increment: Depends on the <Time1> value
*RST:
20E-9
TRIGger:A:WIDTh:DELTa
<Delta>
Defines a variation range around the width value specified using
.
Parameters:
<Delta>
Variation
±Δ
t
Range:
Minimum is 0. Maximum depends on the defined
pulse width (TRIG:A:WIDTH:WITDH).
Trigger
www.allice.de
Allice Messtechnik GmbH