Remote Control Commands
R&S
®
RTM3000
706
User Manual 1335.9090.02 ─ 08
SEARch:PROTocol:ARINc:FORMat
Searches for a data format.
The command is relevant if
SEARch:PROTocol:ARINc:CONDition
LDATa
.
Parameters:
<DataFormat>
DATA | DSSM | DSDI | DSSSm
SSMData
: SSM+Data
SDIData
: SDI+Data
SSData
: SSM+Data +SDI
SEARch:PROTocol:ARINc:LABel:CONDition
Sets the comparison condition for label: If the pattern contains at least one X (don't
care), you can trigger on values equal or not equal to the specified value. If the pattern
contains only 0 and 1, you can also trigger on a range greater than or lower than the
specified value.
The command is relevant if
SEARch:PROTocol:ARINc:CONDition
LABel
or
LDAta
.
Parameters:
<LabelCondition>
EQUal | NEQual | GTHan | GEQual | LEQual | LTHan | WITHin |
OUTSide
SEARch:PROTocol:ARINc:LABel:MAXimum
Searches for a maximum value of the label if
INRange
or
OORange
.
The command is relevant if
SEARch:PROTocol:ARINc:CONDition
LABel
or
LDATa
.
Parameters:
<LabelMaximum>
01X-string
SEARch:PROTocol:ARINc:LABel:MINimum
Searches for a minimum value of the label.
The command is relevant if
SEARch:PROTocol:ARINc:CONDition
LABel
or
LDATa
.
Parameters:
<LabelMinimum>
01X-string
SEARch:PROTocol:ARINc:SDI
Searches for the specified source/destination identifier (SDI) bits.
The command is relevant if
SEARch:PROTocol:ARINc:CONDition
LDATa
.
Serial Bus Analysis
www.allice.de
Allice Messtechnik GmbH