Remote Control Commands
R&S
®
RTM3000
590
User Manual 1335.9090.02 ─ 08
Calibration?
returns information on the state of the self-alignment. Return values ≠
0 indicate an error.
Same as
*CAL?
.
Return values:
<SelfAlignment>
Numeric status indicator
CALibration:STATe?
Returns the overall state of the self-alignment.
Return values:
<SelfAlignmentState> NOALignment | RUN | ERRor | OK | ABORt
NOALignment: no self-aligment was performed. Relevant for
service operations.
RUN: self-aligment is running
ERRor: an error occured.
OK: self-aligment has been performed successfully
ABORt: self-aligment has been cancelled
Usage:
Query only
SYSTem:NAME
Defines an instrument name.
Parameters:
<Name>
String with max. 20 characters
SYSTem:DATE
<Year>,<Month>,<Day>
Specifies the internal date for the instrument.
Parameters:
<Year>
Increment: 1
Default unit: a
<Month>
Range:
1 to 12
Increment: 1
<Day>
Range:
1 to 31
Increment: 1
Default unit: d
Usage:
SCPI confirmed
SYSTem:TIME
<Hour>,<Minute>,<Second>
Specifies the internal time for the instrument.