Serial Bus Analysis
R&S
®
RTM3000
323
User Manual 1335.9090.02 ─ 08
Remote command:
TRIGger:A:MILStd:SADDress:CONDition
Edit Minimum/Edit Maximum ← Sub Address ← Command
Opens the "Edit Minimum"/"Edit Maximum" submenu.
After setting the "Compare" condition, you can enter the value bit-by-bit by setting the
state high, low or do not care for each single bit. Alternatively, you can enter a hexa-
decimal value.
Remote command:
TRIGger:A:MILStd:SADDress:MAXimum
TRIGger:A:MILStd:SADDress:MINimum
Word Count ← Command
Sets the word count.
Compare ← Word Count ← Command
Sets the condition how the decoded value is compared with the defined range.
Remote command:
TRIGger:A:MILStd:WCOunt:CONDition
Minimum ← Word Count ← Command
Sets the minimum of the word count range.
Remote command:
TRIGger:A:MILStd:WCOunt:MINimum
Maximum ← Word Count ← Command
Sets the maximum of the word count range.
Remote command:
TRIGger:A:MILStd:WCOunt:MAXimum
Sub Address ← Command
When "Cmd. Type" > "Mode Code" is selected, the subaddress selection is restricted
to "0", "31" or "0 | 31"
Remote command:
TRIGger:A:MILStd:SADDress:MCADdress
Mode Code ← Command
Selects the type of mode code when "Command Type" > "Mode Code" is selected.
Remote command:
Status
Triggers on a status word that is specified.
.
Status ← Status
Selects a status bit. For each bit, you can select the "State" to be triggered on.