Remote Control Commands
R&S
®
RTM3000
525
User Manual 1335.9090.02 ─ 08
Value
Description
Queries for results
HORizontal
Sets two horizontal cursor lines and measures the
voltages at the two cursor positions and the delta
of the two values.
VERTical
Sets two vertical cursor lines and measures the
time from the trigger point to each cursor, the time
between the two cursors and the frequency calcu-
lated from that time.
HVERtical
Combines the HORizontal cursor and VERTical
cursor measurements. Two horizontal and two ver-
tical cursor lines are set and the voltages and time
from the trigger point are measured at the cursor
positions, as well as the delta of the voltage and
time values.
CURSor<m>:SOURce
<Source>
Defines the source of the cursor measurement.
Suffix:
<m>
.
1, the suffix is irrelevant.
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4 | MA1 | MA2 | MA3 | MA4 | MA5 | RE1 |
RE2 | RE3 | RE4 | XY1 | XY2 | D70 | D158 | D0..15 | SPECtrum |
MINHold | MAXHold | AVERage
CH1 | CH2 | CH3 | CH4
Active analog channel waveforms 1 to 4
MA1 | MA2 | MA3 | MA4 | MA5
Active math waveforms 1 to 5
RE1 | RE2 | RE3 | RE4
Active reference waveform 1 to 4
XY1 | XY2
Active XY-waveforms
D0..15
Active digital channels D0 to D15, available if MSO option
R&S
RTM-B1 is installed.
On individual digital channels, only vertical (time) cursor mea-
surements are possible.
D70 | D158
Active digital channels D0...D7 (pod
1) and D8...D15 (pod
2).
Only available, if MSO option R&S
RTM-B1 is installed. On
pods, only V-marker measurement is possible.
Measurements