Remote Control Commands
R&S
®
RTM3000
459
User Manual 1335.9090.02 ─ 08
TRIGger:B:ENABle
<State>
Activates the B-trigger.
Parameters:
<State>
ON | OFF
*RST:
OFF
TRIGger:B:SOURce
<Source>
Sets the source of the B-trigger.
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4 | D0 ... D15
CH1 | CH2 | CH3 | CH4
Analog input channels. Available channels depend on the instru-
ment type.
D0..D15
Digital channels D0 to D15. Require MSO option R&S
RTM-B1.
*RST:
CH1
TRIGger:B:EDGE:SLOPe
<Slope>
Sets the slope of the B-trigger.
Parameters:
<Slope>
POSitive | NEGative | EITHer
*RST:
POSitive
TRIGger:B:MODE
<Mode>
Sets a time delay or event delay for the B-trigger. The instrument waits for that delay
after an A-trigger until it recognizes the B-trigger.
Parameters:
<Mode>
DELay | EVENts
DELay
Time delay. Set the time with
EVENts
Event delay. Set the number of events with
*RST:
DELay
TRIGger:B:DELay
<DelayTime>
Sets the time the instrument waits until it checks the B-trigger condition.