Applications
R&S
®
RTM3000
184
User Manual 1335.9090.02 ─ 08
Remote command:
SPECtrum:MARKer:SETup:DISTance
8.3.3.3
R-Marker Settings
Access: "FFT" menu > enable "Peak List" > "Marker" > "R-Marker"
You can define a reference marker and get the peak results as delta values to the val-
ues of the reference marker. The reference marker is displayed on the waveform as a
filled orange triangle.
R-Marker
Defines the mode for the selection of the reference peak.
"Off"
Disables the reference marker.
"Indicated"
The reference marker is set to the peak with a specified index num-
ber.
"Range"
The peak with the highest level within the specified range is set as
the reference marker.
Remote command:
SPECtrum:MARKer:REFerence:SETup:MODE
Index
Defines the number of the peak to which the reference marker is set if "R-Marker" =
"Indicated"
Remote command:
SPECtrum:MARKer:REFerence:SETup:INDex
Frequency
Defines the center frequency for the capture range if "R-Marker" = "Range".
Spectrum Analysis and Spectrogram (Option R&S
RTM-K18)