The instrument captures and decodes the signal according to the protocol definition
and the configuration settings.
The color-coding of the various protocol sections and errors simplifies the interpretation
of the visual display. The decode information condenses or expands, depending on the
horizontal scale. Various data formats are available to show the result values.
Figure 12-33: Decoded LIN signal with bus table, trigger on frame start
Table 12-5: Content of the LIN frame table
Column
Description
Start time
Time of frame start in relation to the trigger point
ID
Identifier value, hexadecimal value
Length
Number of data bytes
Data
Hexadecimal values of the data bytes
Chks
Checksum value
State
Overall state of the frame.
Remote commands are described in
Chapter 16.11.6.3, "LIN - Decode Results"
12.6.5
Search on Decoded LIN Data
Using the search functionality, you can find the same events in the decoded data which
you also can trigger on. Unlike trigger, the search finds all events in an acquisition that
fulfill the search condition. The results are listed in a table and and can be saved to file.