H600/H605 R
EPAIR
& R
EPLACEMENT
- P
AGE
25
1041895, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
7. Remove the O
2
/solenoid/battery wiring harness connector from the J1 location on the Main PCB.
F
IGURE
AA: P
OWER
S
WITCH
AND
M
AIN
PCB J1 L
OCATION
8. Remove the screws that secure the screw down cable ties to the Front Cabinet.
9. Either feed the O
2
/solenoid/battery wiring harness through the screw down cable ties or, if new
screw down cable ties are available, cut the cable ties and remove the O
2
/solenoid/battery wiring
harness.
10. Note the location at which the battery wires exit the large convoluted tubing. Cut the cable ties and
remove the O
2