M600/M605 R
EPAIR
& R
EPLACEMENT
- P
AGE
31
1041896, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
M
OTOR
S
TART
/R
UN
C
APACITOR
To remove the Motor Start/Run Capacitor
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove the Compressor Cover. Refer to the Compressor Cover Replacement Section.
F
IGURE
AE: C
APACITOR
L
OCATION
4. Using insulated needle-nose pliers, carefully remove the female connectors on the Capacitor wires
from the two Capacitor terminals.
5. After the Capacitor is discharged, lift the Capacitor up and out of the molded base.
To install the Motor Start/Run Capacitor
1. Slide the Capacitor into position between the foam insulation and the Compressor assembly
standoff.
2. Connect the Capacitor wires to the terminals on the Capacitor.
3. Make sure the Capacitor terminals are parallel to the rear wall in the Base Cabinet.
CAUTION
Short the terminals of the Capacitor to discharge.