M
AINTENANCE
- P
AGE
23
1041813, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
24. Connect the T-27 Trox bit to the socket wrench. Press down on the wrench and turn counter
clockwise to loosen the screw. If the screw does not start to turn or the bit begins to slip, reapply
heat for one minute. Continue this process until the screw is removed. Once the parts have
cooled, discard the piston cup and screw.
F
IGURE
AC: S
CREW
R
EMOVAL
25. Place a new cylinder sleeve over the piston head. Place a new piston cup on the back of the cup
retainer plate. Note the locations of the alignment pin on the back of the plate and the alignment
hole on the piston head.
F
IGURE
AD: R
EPLACE
P
ISTON
C
UP
26. Press the plate into the cylinder sleeve being careful not to crimp the piston cup. Once the plate is
seated on top of the piston, rotate the plate until the plate pin seats into the piston alignment hole.
Tighten the new T-27 plate screw to 100 in-lbs. Remove the rod assembly from the vice.
27. With the eccentric side of the connecting rod/eccentric assembly facing away from the Compres-
sor, slide the assembly onto the shaft until the assembly is flush against the motor assembly.