H600/H605 R
EPAIR
& R
EPLACEMENT
- P
AGE
47
1041895, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
P
RESSURE
R
EGULATOR
R
EPLACEMENT
To remove the Pressure Regulator
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove the pressure tubing from the outlet port on the pressure regulator.
4. Note the orientation of the pressure regulator outlet port. The new pressure regulator must be
installed with the outlet port in the same position.
5. Using a pair of channel locks, remove the pressure regulator from the Sieve Canister Assembly.
F
IGURE
AU: R
EMOVING
THE
P
RESSURE
R
EGULATOR
To install the Pressure Regulator
1. Apply a layer of Teflon thread tape to the threads of the pressure regulator.
2. Align the pressure regulator with the pressure regulator port in the Sieve Canister Assembly.
Screw the pressure regulator into the Sieve Canister Assembly until strong resistance is felt and
the pressure regulator outlet port is oriented in the correct position.
3. Install the pressure tubing on the pressure regulator outlet port.
4. Connect the unit to a power source and turn on the unit. Apply snoop leak detector to the area
where the pressure regulator mounts to the Sieve Canister Assembly and where the O
2
tubing
connects to the outlet barb to check for leaks.
CAUTION
The tape should be installed in a counter clockwise direction, with the Pressure
Regulator threads facing you.