M10600/M10605 R
EPAIR
& R
EPLACEMENT
- P
AGE
43
1041898, R
EV
. 01
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
B
ASE
C
ABINET
R
EPLACEMENT
To Remove the Base Cabinet
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove Compressor Cover/Perforated Canopy. Refer to the Compressor Cover Replacement
Section
4. Remove Capacitor. Refer to the Motor Run/Start Capacitor Replacement Section.
5. Remove Compressor/motor assembly. Refer to Compressor/Motor Assembly Replacement Sec-
tion.
6. Remove In-line Muffler. Refer to In-line Muffler Replacement Section.
7. Remove Hour Meter. Refer to Hour Meter Replacement Section.
8. Remove Sieve Canister Assembly. Refer to Sieve Canister Replacement Section.
9. Remove Power Cord. Refer to Power Cord Replacement Section.
To Install the Base Cabinet
1. All components have now been removed from the original Base Cabinet. To begin reassembly,
place the Base Cabinet on a protected work surface. Follow the disassembly process in reverse
order to reassemble the unit.
NOTE
If care is taken during the following Section, the Valve/Solenoid Assembly and the
pressure regulator do not have to be removed. However, if these components are
damaged during the procedure, they must be replaced.