M10600/M10605 R
EPAIR
& R
EPLACEMENT
- P
AGE
13
1041898, R
EV
. 01
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
To install the Power Switch
1. Orient the Power Switch so that the single terminal is at the top. Align the Power Switch with the
opening in the Front Cabinet assembly. Press the Power Switch into the Front Cabinet assembly
until all four latches on the Power Switch lock in place.
2. Install the wires onto the terminals of the Power Switch. Ensure that they are installed in their orig-
inal position.