Power is supplied from the two-prong plug to the Power Switch. When the Power Switch is ON, 120 VAC/230
VAC power is supplied to the Compressor motor, Hour Meter, cooling fan, and the main circuit board.
The main circuit board consists of a linear power supply, pressure sensor, software driven microprocessor,
green, yellow, and red LED’s and audible alarm.
The 9-volt battery provides a power source to operate the audible alarm and the red LED alarm light in the
event of an AC power failure.