T
ESTING
- P
AGE
11
1041903, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
M
ODEL
600
W
/P
EDIATRIC
F
LOW
M
ETER
T
EST
D
ATA
S
HEET
NOTE
All information on this data sheet should be entered in the correct location after the associated test was
completed. The data sheet must then be signed in ink and dated by the technician performing tests.
Step 2
Serial Number: __________
Step 3
VAC: __________ Hz:__________
Step 4
Hour Meter: __________
NOTE
Step 5 is not required for this Model. The Alarm will not
sound with the 1 Liter Pediatric Flowmeter installed.
Step 6
Oxygen Outlet Pressure
5.0-7.0 psig
PRESSURE GAUGE READING
NOTE
Step 7 is not required for this Model.
Step 8
Flow Meter
Oxygen
Concentration
H600; M600; Enhanced M600
120V
> 92% @ 0.1 lpm to 1 lpm
0.1 lpm
1 lpm
Signature: ____________________ Date:_______________ Notification Number: _____________