To install the Flow Meter
1. Align the threaded fittings on the back of the supplied Flow Meter with the holes in the Front Cabi-
net.
2. While holding the Flow Meter in place, install the support speed nuts to secure the Flow Meter to
the Front Cabinet. Hand tighten the speed nuts.
3. Connect the pressure tubing to the fittings on the Flow Meter, then secure the tubing with the cable
ties.
4. Turn on the unit, and check the flow tube connections for leaks.
5. If no leaks are detected, turn off the unit, reassemble, and test.
NOTE
The Flow Meter must be mounted with the shaft for the Flow Meter knob at the
top.