E
NHANCED
M600/605; E
NHANCED
M10600/M10605 R
EPAIR
& R
EPLACEMENT
- P
AGE
41
1041897, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
P
RESSURE
R
EGULATOR
R
EPLACEMENT
To remove the Pressure Regulator
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove the pressure tubing from the outlet port on the pressure regulator.
4. Note the orientation of the pressure regulator outlet port. The new pressure regulator must be
installed with the outlet port in the same position.
F
IGURE
AJ: O
RIENTATION
OF
THE
P
RESSURE
R
EGULATOR
5. Using a pair of channel locks, remove the pressure regulator from the Sieve Canister Assembly.
F
IGURE
AK: R
EMOVAL
OF
THE
P
RESSURE
R
EGULATOR