1041813, R
EV
. 00
P
AGE
4 - M
AINTENANCE
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
C. C
LEANING
THE
C
ABINET
1. Disconnect the unit from the power source.
2. Using a mild detergent and disinfectant, clean all exterior surfaces of the cabinet.
3. Ensure that the cabinet is dry before placing in service.
D. M
ICRO
-
DISK
F
ILTER
M
AINTENANCE
The Micro-disk filter is the final component in the filtering system before the concentrated oxygen is delivered
to the patient. Check and replace this filter as necessary.
Procedure
1. Remove the Rear Cabinet.
CAUTION
Do not allow any liquid to enter the unit.
PART NUMBER: H621
Included in Kit
Tools Required
Micro-disk Filter (x6)
Wire Ties (x2)
Phillips Screwdriver
(No. 2 Medium w/ long shaft)
Diagonals (wire cutters)
WARNING
Electrical shock hazard; Disconnect the electrical supply before attempting to make
any repairs.
CAUTION
Electronic components used in this device are subject to damage from static
electricity. Repairs made to this device must be performed only in an antistatic,
ESD protected environment.