1041897, R
EV
. 00
P
AGE
32 - E
NHANCED
M600/605; E
NHANCED
M10600/M10605 R
EPAIR
& R
EPLACEMENT
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
To Install the Main PCB to Compressor Wiring
1. Connect the white wire from the Main PCB to Compressor wiring harness to the brown wire com-
ing from the Compressor. Ensure that the wire connectors are fully seated. Refer to the block dia-
gram in Figure AA.
2. Connect the black wire from the Main PCB to Compressor wire harness to the blue wire coming
from the thermal protection switch. Ensure that the wire connectors are fully seated. Refer to the
block diagram in Figure AA.
3. Connect the white wire from the Main PCB to Compressor wire harness to terminal J6 on the Main
PCB assembly. Ensure that the wire connectors are fully seated. Refer to the block diagram in
Figure AA.
4. Connect the black wire from the Main PCB to Compressor wire harness to terminal J20 on the
Main PCB assembly. Ensure that the wire connectors are fully seated. Refer to the block diagram
in Figure AA.