1041897, R
EV
. 00
P
AGE
44 - E
NHANCED
M600/605; E
NHANCED
M10600/M10605 R
EPAIR
& R
EPLACEMENT
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
To install the Sieve Canister Assembly
1. Install the supplied canister foam on the bottom of the Sieve Canister Assembly.
2. Install the brass fitting in the top of the Sieve Canister Assembly.
3. Install the Blow Down Muffler onto the Sieve Canister Assembly. Refer to the Blow Down Muffler
Replacement Section.
4. Set the Sieve Canister Assembly in place in the Base Cabinet.
5. Slide the one-eared clamp provided onto the 12” pressure tubing.
6. Install the 12” pressure tubing provided on the inlet port of the Sieve Canister.
7. Using a crimping clamp tool or diagonals, “pinch” the one-eared clamp securing the pressure tub-
ing to the inlet port of the Sieve Canister Assembly.
8. Using the 5/16” Nut Driver, insert then tighten the four screws that secure the Sieve Canister
Assembly to the molded base.
9. Install the pressure regulator. Refer to the Pressure Regulator Replacement Section.
10. Install the Valve/Solenoid on the top of the Sieve Canister. Refer to the Valve/Solenoid Replace-
ment Section.
11. Install the wire connectors on the terminals of the valve/solenoid.
12. Connect the yellow pressure tubing to the brass fitting on the top of the Sieve Canister.
13. Connect the clear pressure tubing to the port on the pressure regulator.
14. Install the Compressor Cover. Refer to the Compressor Cover Replacement Section.
NOTE
Ensure that the AC Power Cord is clear of the Sieve Canister Assembly.